Application Bulletin 356

Installation instruction for «VoltIC pro I»

Page 7 of 18

2.5 Connection of devices

The 858 Professional Sample Processor, the 850

Professional IC AnCat system and the 797 VA Computrace

are connected with each other via the 800 Dosino with a 10

mL Dosing Unit mounted on the 858 Professional Sample

Processor.

Connect the sample needle to Port 3 of the 10 mL Dosing

Unit at the 800 Dosino using a PEEK capillary (70 cm,

d=0.75 mm, 6.1831.070). The 850 Professional IC AnCat

system is connected to the Dosing Unit at Port 1 using a

PTFE capillary (90 cm, d=0.25 mm, part of 6.1803.150). The

sample inlet capillary (PEEK, 90 cm, d=0.5 mm, part of

6.1822.410) of the 797 VA Computrace is connected to the

Dosing Unit at Port 2. Port 4 of the Dosing Unit is equipped

with an adapter (6.1808.280) and a FEP tube (6.1805.530)

and connected to the waste.

The second 10 mL Dosing Unit with an 800 Dosino is

connected to one capillary of the fourfold micro tip and is

used to dilute the sample for the VA measurement

automatically (solution j).

Fig.: Capillary connections to and from rinsing station