Installation of 797 va computrace software, Configuration of 797 va computrace software, Starting a determination – Metrohm VoltIC pro I User Manual

Page 13

Advertising
background image

Application Bulletin 356

Installation instruction for «VoltIC pro I»

Page 13 of 18

Metrosep A Supp 5 - 150/4.0 to the analysis anion, assign

the column Metrosep C 4 - 100/4.0 to the analysis cation,

assign the two eluents to their respective IC pump, match

the two suppressor solutions to the IC peristaltic pump, and

define the rack of the connected sample processor. Adjust

the temperature of the column oven to 30°C. Save the

method.

In order to start the equilibration, go to the window work

place, load your method, and press «Start HW». Before

inserting the columns, flush the system for about 10 minutes

and get rid of air bubbles (by using the purge valve and

syringe), then insert and rinse the precolumns for 10

minutes, and only afterwards put the analytical columns into

the flow (see also 850 manual).

3.4

Installation of 797 VA Computrace software

Install the software. All the standard directories proposed by

the program should be accepted. Restart windows.

Connected USB devices are automatically recognized when

VA Computrace software is started.

3.5

Configuration of 797 VA Computrace software

Make the following general settings in the main window

under «Settings»  «General Settings»

s” and click «OK»:


Load method «Det of Zn Cd Pb Cu_IC_VA.mth», modify or

create a new method.

Click on «Dosinos» in the «Working Method Specifications»

window and make following settings: