2 Installation

837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003

14

2.4 Connecting

to

the 761 Compact IC

The installation of the 837 IC (Eluent/Sample/Combi) Degasser with the

761 Compact IC, and its control with the «761 PC Software» software is

described below.

2.4.1 Electrical

connection to the 761 Compact IC

Always switch off the 837 IC (Eluent/Sample/Combi) Degasser before

making new electrical connections.


The electrical connections of the system consisting of the 761 Compact

IC and the 837 IC (Eluent/Sample/Combi) Degasser are made accord-

ing to the following figure. To activate the communication between the

761 Compact IC and 837 IC (Eluent/Sample/Combi) Degasser, the re-

mote-switch (10) on the Degasser must be set to ON.

761

837

6.2143.210

6.2134.100

PC

Fig. 6:

Connecting the 837 IC (Eluent/Sample/Combi) Degasser
to the 761 Compact IC

The remote-cable 6.2143.210 has a two-headed end with two DIN-

plugs. The end C (labeled at the end of the cable) should be used for

the Degasser. The end B is only needed, if an 828 IC Dual Suppressor

is connected simultaneously.

2.4.2 Settings

for

control

with the «761 PC Software»

Is the 837 IC (Eluent/Sample/Combi) Degasser connected to the 761

Compact IC via end C of the remote-cable 6.2143.210 as described

above, the value for Remote Line 2 in the software «761 PC Software»

must be set to 1 (see Instructions for Use 761 Compact IC).