4 Troubleshooting - Problems

837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003

18

4

Troubleshooting - Problems

4.1

Remedying faults and problems

If difficulties occur during analyses with your IC system then it is best to

search for their causes in the following sequence: column

pump

eluent

IC system. In the "Instructions for Use" of your Modular IC

system or 761 Compact IC you will find an overview of possible faults

together with their causes and remedies.
In addition to these general problems, the following section covers

those problems which could arise from the use of the 837 IC (Elu-

ent/Sample/Combi) Degasser.

4.2 Chromatography

problems

Instrument

Problem

Cause

Remedy

Some peaks lar-
ger/smaller
than ex-
pected

Carry-over of sample

from the previous mea-
surement

Extend rinsing time between two

samples, see section 4.4

Sample/Combi Degas-
ser (Sample flow path
used)

2.837.0020
2.837.0030

Peaks smaller than
expected

Leak in system (Sample
flow path)

Blockage in the flow

path (Sample flow path)

Check if liquid is leaking from
the exhaust opening 15. If yes,

contact Metrohm Service.

Contact Metrohm Service

Poor retention time
reproducibility

Leak in system (Eluent
flow path)

Blockage in the flow

path (Eluent flow path)

Check if liquid is leaking from
the exhaust opening 15. If yes,

contact Metrohm Service.

Siehe Kap.4.3.

Eluent/Combi Degas-
ser (Eluent flow path
used)

2.837.0010
2.837.0030

Noisy or unstable

baseline

Leak in system (Eluent

flow path)

Blockage in the flow
path (Eluent flow path)

Check if liquid is leaking from
the exhaust opening 15. If yes,

contact Metrohm Service.

Contact Metrohm Service

4.3 Instrument

problems

Problem

Cause

Remedy

LED "Power" 1 does not
light up (Mains switch 12 on

I )

Mains cable not connected

Fuse blown

Electronics fault

Connect the Degasser to the mains sup-

ply

Replace fuse, see section 2.5.2

Contact Metrohm Service