5 mains connection, 1 mains voltage, 2 fuse – Metrohm 837 IC Degasser User Manual

Page 21: Mains connection, Mains voltage, Fuse

Advertising
background image

2.5 Mains connection

837 IC (Eluent/Sample/Combi) Degasser/ Instructions for Use 8.837.1003

15

2.5 Mains

connection

Please observe the following rules when connecting the instrument to

the electricity supply. If the instrument is operated with an incorrectly

set mains voltage and/or an incorrect mains fuse then it represents a

fire hazard!

2.5.1 Mains

voltage

All power supplies from 100 V to 240 V with 50/60 Hz can be connected

to the mains input of the 837 IC (Eluent/Sample/Combi) Degasser.

Adaptation to the mains supply used is carried out automatically.

2.5.2 Fuse

A 1.0 A fuse (slow blow) is contained in the fuse holder (11) of the 837

IC (Eluent/Sample/Combi) Degasser.

Make sure that the instrument is never operated with a different type of

fuse as otherwise it represents a fire hazard!

Changing the fuse

1 Pull out mains cable

Switch off the instrument and remove the mains cable from

mains supply connection (13) of the 837 IC (Eluent/Samp-

le/Combi) Degasser.

2 Remove the fuse holder

Use a screwdriver to loosen the fuse holder beside the mains

supply connection (see Fig. 4) and remove it completely.

3 Changing the fuse

Change the fuse if necessary and replace it in the fuse holder.

Use only a fuse with the following specifications:

1.0 A fuse (slow blow) Metrohm-No. U.600.0016

4 Re-insert fuse holder

Advertising