10 peristaltic pump, 11 metrohm suppressor module (msm), 12 mains connection – Metrohm 881 Compact IC pro – Anion User Manual

Page 108: 13 interfaces

Advertising
background image

7.10 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

98

■■■■■■■■

881 Compact IC pro – Anion

7.10

Peristaltic pump

Type

2-channel peristaltic pump

Rotating direction

Counterclockwise/Clockwise rotation

Rotational speed

0…42 rpm in 7 stages at 6 rpm.

Pumping proper-
ties

0.3 mL/min at 18 rpm; with 6.1826.320 standard pump tubing.

Material of pump
tubings

recommended: Tygon Long Flex Life

7.11

Metrohm Suppressor Module (MSM)

Resistance to sol-
vents

100 % no restriction

Switching time

typ.100 ms

7.12

Mains connection

Required voltage

100…240 V ± 10 % (autosensing)

Required fre-
quency

50…60 Hz ± 3 Hz (autosensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

internal fuse 3.15 A

7.13

Interfaces

USB

Input

1 USB upstream, type B (for connection to the PC)

Output

2 USB downstream, type A

MSB

2 MSB Mini-DIN 8-pin (female) (for Dosino, stirrer, remote lines, …)

Caution

When connecting an instrument to the MSB connector you must
switch off the 881 Compact IC pro.