3 drainage tubings, Drainage tubings, Figure 5 – Metrohm 881 Compact IC pro – Anion User Manual

Page 31: Leak sensor – plugging in, 3) from the base tray, Plug the leak sensor connector plug (5-2), Nection socket (5-1)

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

881 Compact IC pro – Anion

■■■■■■■■

21

1 2

3

Figure 5

leak sensor – plugging in

1

Leak sensor connection socket
Is labeled with Leak Sensor.

2

Leak sensor connector plug

3

Leak sensor connection cable
Extractable. Coiled up in the base tray.

3.6.3

Drainage tubings

Fluid that escapes in the covering plate or in the detector chamber flows
through the drainage tubings into the base tray and past the leak sensor
into the waste container. This ensures that any leaks in the system will be
detected by the leak sensor.

Advertising