3.14 Column heater
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
881 Compact IC pro – Anion
3.13.3
Selecting the sample loop
The amount of sample solution injected depends on the volume of the
sample loop. The choice is made on the basis of the application. The fol-
lowing sample loops are normally used:
Cation determination
10 µL
Anion determination with suppression
20 µL
Anion determination without suppression
100 µL
3.14
Column heater
The perfect isolation of the column chamber ensures thermally stable con-
ditions for the separation column. The temperature of the column heater
can be set in the software.