Course of events, Brightener determination series, Manual operation – Metrohm MVA-20 User Manual

Page 11: Workplace

Advertising
background image

Application Bulletin 402/1 e

Installation instruction for MVA-20

5.1.3. Course of events

The execution of calibration and determination by

dilution titration is controlled by the 894 Professional

CVS and viva.

a) The measuring vessel is emptied using the 843

Pump Station.

b) Dosing Unit «894/D4 2mL Std or sample»

connected to Dosino 4 is automatically rinsed and

prepared with suppressor standard or sample from

the vial on the rack of the sample changer.

c) The measuring vessel is rinsed using the 843

Pump Station.

d) Dosino 1 automatically adds the VMS and Dosino

4 is used for the automatic addition of suppressor

standard or plating bath sample.

e) At the end of each determination the measuring

vessel, is automatically emptied and rinsed using

the 843 Pump Station.

Next sample starts again from a).

Recalibration within a determination series is possible.

At the end of a determination series Dosino 4 is

automatically prepared with rinsing solution.

5.2. Brightener determination series

5.2.1. Manual operation

Dosing Unit «894/D1 50mL VMS» connected to Dosino

1 has to be prepared with VMS.

Dosing Unit «894/D2 2mL Brightener» connected to

Dosino 2 has to be prepared with brightener

concentrate.

Dosing Unit «894/D3 2mL Suppressor» connected to

Dosino 3 has to be prepared with suppressor

concentrate.

Plating bath samples are placed on position 1 to 14 of

the sample rack using the sample vials for 50 mL

(6.2747.010). The exact sample volume needed for the

determination has to be pipetted into the vial.

Rinsing solutions are placed on position 15 to 28 of the

sample rack using the sample vials for 50 mL

(6.2747.010). Usually the same volume of deionized

water is used for rinsing like it was used for the sample.

The rinsing solution has to be placed diagonally

opposite the sample on the second half of the circle

(Sample position + 14).

Fig. 16: Example positions of solutions on the sample rack
6.2041.490 for brightener determination

5.2.2. Workplace

In the «Run» window on the tab «Determination series»

create a sample table

Fig. 17: Example sample table for brightener determination

As sample type S

AMPLE

has to be selected.

For «Sample amount» the volume pipetted into the vial

has to be entered.

5.2.3. Course of events

The execution of the determination by modified linear

approximation technique is controlled by the 894

Professional CVS and viva.

a) The measuring vessel is emptied using the 843

Pump Station.

b) The intercept solution is automatically prepared at

the beginning of each determination, by dosing

Page 11 of 12