Rack, Rack data, Needle adjustment – Metrohm MVA-20 User Manual

Page 7

Advertising
background image

Application Bulletin 402/1 e

Installation instruction for MVA-20

3.5.2. Rack

Rack name

6.2041.490

Shift rate

20

°/s

3.5.3. Rack data

Rack parameters

Beaker radius samples

off

mm

Beaker sensor

off

Rack offset

0

°

Lift positions – Tower 1

Work position

User

defined, see

3.5.4

mm

Rinse position

0

mm

Shift position

0

mm

Special position

0

mm

Lift positions – Tower 2

Work position

Not available

Rinse position

Not available

Shift position

Not available

Special position

Not available

Special beakers

Not used

3.5.4. Needle adjustment

The work position of the pipetting needle 6.1835.050 has to

be adjusted in a way that the lower end of the needle is

positioned max. 1 mm above the bottom of the sample vial.

This is essential to guarantee a complete transfer of the

sample from the sample vial on the rack of the 919 IC

Autosampler plus into the measuring vessel of the 894

Professional CVS.

Fig. 9: Needle adjustment

Procedure to adjust the needle position in viva

1. Insert an empty sample vial to one of the sample

positions on the rack (position 1 to 28).

2. Click on the Manual symbol.

3. Select Tower 1 (Sample changers – 919 IC

Autosampler plus 1).

4. Open tab «Move».

5. Enter the position of the empty sample vial in the field

«Target position».

6. Click Start to move the needle to the chosen rack

position.

7. Enter 125 mm in the field «Target position» in the

section «Lift position».

8. Click Start to lower the needle to the chosen target

position. For the exact positioning of the needle as

shown in «Fig. 9», use the arrow up and down buttons

to adjust the needle position stepwise.

9. Once the needle is properly adjusted select the tab

«Assign position».

0.5 … 1 mm

Page 7 of 12