Viva «method, Method run, Measuring commands – Metrohm MVA-20 User Manual

Page 8: Dosing commands, Automation commands, Transferring sample

Advertising
background image

Application Bulletin 402/1 e

Installation instruction for MVA-20

10. The new value can be seen in the section «Lift position

– Current position».

11. Select the option «Work position for Tower» in the

section Lift position.

12. Click on «Assign» in the section Lift position.

13. Leave the Manual control with a click on the «Close»

button.

4. viva «Method»

4.1.

Method run

viva software includes three method templates for CVS

analysis with an automated system.

Brightener determination (CVS, MLAT), automated

Conditioning (Cu VMS), automated

Suppressor determination (CVS, DT), automated

The templates already establish the basic sequences for the

different applications, the use of 800 Dosinos for automatic

dosing as well as the calculation of the result and the

application of a sample changer and pumps to process a

sample series. However certain commands and settings

have to be adapted to the used hardware and the

requirements of the specific application.

For a step by step description please see the Tutorial

chapter «6.2 Methods for the automated determination».

The following commands have to be adapted:

4.1.1. Measuring commands

Such as

Assign the 894 Professional CVS instrument to be used

on the tab «General/Hardware».

Adapt the measuring parameters on the tabs «Pre-

treatment», «Sweep», «Post-treatment» and

«Potentiostat» according to the requirements of the

application. These parameters can be found in separate

application documentation.

4.1.2. Dosing commands

Such as:

Define the volume of solution that should be dosed

when this command is due.

Such as:

Define the volume of solution that should be dosed

when this command is due.

Assign the Dosing Unit that should be used to run this

command.

Such as:

On the tab «General/Hardware» assign the 894

Professional CVS instrument to which the 800 Dosino is

connected.

4.1.3. Automation commands

Such as:

Assign the 919 IC Autosampler plus that should be

used to run this command.

4.1.4. Transferring sample

This only applies for the brightener determination by MLAT!

To guarantee a complete transfer of the sample from the

sample vial on the rack of the 919 IC Autosampler plus into

the measuring vessel of the 894 Professional CVS the

peristaltic pump time has to be adjusted in the TRACK

P

ERISTALTIC PUMP

in the

command PUMP

P

ERISTALTIC

PUMP

ON.

Fig. 10: Method snippet TRACK

P

ERISTALTIC PUMP

Page 8 of 12