Application Bulletin 402/1 e

Installation instruction for MVA-20

4.1.5. Rinsing measuring vessel

After each sample the measuring vessel need to be rinsed

with deionized water. The rinsing and draining times depend

on the total volume used in the determination.

If the method templates are used, the rinsing and draining

times are defined in the corresponding PUMP

commands in

the TRACK

R

INSE

and

TRACK

D

RAIN

. The number of

rinsing cycles is defined in the command LOOP

LOOP

R

INSING

in the TRACK

R

INSING CYCLES

.

Fig. 11: Method snippet TRACK

R

INSING CYCLE

Examples for rinsing and draining times:

Cell volume

50 mL

100 mL

Number of rinsing cycles

2

2

Draining time [s]

35

50

Rinsing time [s]

9

18

Please note! The device assigned to the commands PUMP

PUMP

R

INSE

und PUPM

PUMP

D

RAIN

need to be the

919 IC Autosampler plus, since the 843 Pump Station is

connected to and controlled from the sample changer.

4.1.6. Rinsing transfer tubing

This only applies for the brightener determination by MLAT!

After each sample the transfer tubing from the 919 IC

Autosampler plus to the measuring vessel of the 894

Professional CVS need to be rinsed with deionized water.

The rinsing solution is placed in a vial on the rack of the 919

IC Autosampler plus. The position of the rinsing solution

relative to the sample position is defined in the TRACK

G

O

TO RINSING POSITION

in the command MOVE

MOVE

R

INSING

POSITION

.

Fig. 12 Method snippet TRACK

G

O TO RINSING POSITION

To change the relative position of the rinsing solution the

formula editor has to be used on the parameter «Number»

in the section «Target» in the command MOVE

MOVE

R

INSING POSITION

.

Fig. 13 Formula editor to change the rinsing position

The transfer time for the rinsing solution corresponds to the

sample transfer time defined in the TRACK

P

ERISTALTIC

PUMP

in the

command PUMP

P

ERISTALTIC PUMP

ON.

4.2.

Evaluation

Settings regarding evaluation and documentation of the

determination are located in the «Evaluation» part of the

method. The templates already include all necessary

settings to determine brightener or suppressor concentration

in an acid copper bath. If modifications however should be

necessary, here is where important parameters are found: