4 Installation
940 Professional IC Vario ONE/SeS/Prep 3 (2.940.1430)
■■■■■■■■
25
CO
2
adsorption cartridge, an H
2
O adsorption cartridge (6.2837.010) is
also installed.
Accessories
For this step you need the following accessories:
■
Adsorption cartridge holder (6.2057.080) with 4 clips, an adapter
(6.1808.190) and a piece of PVC tubing.
■
CO
2
adsorption cartridge (6.2837.000)
■
H
2
O adsorption cartridge (6.2837.010)
All of the parts are located in one accessory kit: Vario/Flex SeS
(6.5000.020).
Installing the adsorption cartridge
CAUTION
The following preparatory steps absolutely must be carried out for CO
2
suppression to operate correctly.
Preparing the adsorption cartridges:
■
Prepare the H
2
O adsorption cartridge for use as specified in the provi-
ded leaflet.
■
Remove the protective cap on the tip of the CO
2
adsorption cartridge.