Metrohm 940 Professional IC Vario ONE/SeS/Prep 3 User Manual

Page 46

Advertising
background image

4.22 Connecting the separation column

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

940 Professional IC Vario ONE/SeS/Prep 3 (2.940.1430)

Rinse the separation column as follows.

Connecting the separation column

1 Rinsing the separation column

Place a beaker under the separation column's outlet.

Start manual control in MagIC Net and select the high-pressure
pump: Manual

Manual control Pump

Flow: in accordance with column leaflet
On

Advertising