Metrohm 940 Professional IC Vario ONE/SeS/Prep 3 User Manual
Page 39
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
940 Professional IC Vario ONE/SeS/Prep 3 (2.940.1430)
■■■■■■■■
31
Connecting the capillaries of the SMP
NOTE
Pump tubing connections with a filter (6.2744.180) have to be installed
on the outlet of the peristaltic pump's pump tubing in order to protect
the SPM from foreign particles and bacterial growth.
CAUTION
As the PTFE capillaries are very soft, the pressure screws should not be
overtightened.
Squeezed capillary ends can be shortened with the capillary cutter
(6.2621.080).
Connect the SPM capil-
lary …
… to …
… using this accessory
in
Sample Processor
or
(see Sample Processor
manual)
or
a 6-port valve used for
measuring the sample
quantity.
6.2744.010
out
Sample input of the injec-
tion valve
6.2744.010
rinsing solution
Pump tubing output for
rinsing solution
Pressure screw
(6.2744.070)
regenerant
Pump tubing output for
regeneration solution
Pressure screw
(6.2744.070)
waste rins.
Waste container
Directly or via waste collec-
tor (6.5336.000)
waste reg.
Waste container
Directly or via waste collec-
tor (6.5336.000)