Running samples

8.1

Creating a sample table and starting the analysis

1.

Under Run Samples, go to Sample Queue.

The instrument status must be IDLE before starting a measuring series.

2.

Start filling in the sample table.

Start with a line for equilibration of the system.
Use the Equilibrate function for this.

Enter a long run time (e.g. 60 min).
This makes sure that the system has a stable baseline before moving on to the next line.
If the baseline was stable before reaching the total run time, it would still be possible to
move to the next line immediately (with Abort > Move to Next Line).