11 restrictions, Restrictions – Metrohm MagIC Net Empower Driver User Manual

Page 52

Advertising
background image

50

▪▪▪▪▪▪▪

11

Restrictions

This device driver allows you to operate one or several IC systems from one PC. One IC
system may include a maximum of one changer, five Dosinos and one IC instrument
with a conductivity detector.

Two Metrohm IC instruments can be operated in parallel in two different systems.

Two sample changers cannot be operated (with one PC) with this version.

The stop all command does not only stop all components, but also the communication.
As a result, the values are no longer updated in the status display.

The column thermostat remains switched on with the Abort Run command.

The MSM autostep is continued if the sample table is canceled with the Abort after
vial is completed
function. The column thermostat remains switched on as well. All
other components are stopped.

Information on the column used can be obtained via the post run report.

The needle of the sample changer can only move to positions on the rack used (no
external positions such as the rinsing station).

If a Prep or Empty command is performed with the Dosino, then the "Dosing port 1"
will always be used as the port for ejecting the solution (Port for Prep/Empty).

For cation determination, the peaks in the live display are negative. In the analysis
results, the chromatograms are displayed correctly (with positive peaks). For anion
determination, the peaks are always positive.

If the analytes are measured as negative peaks, the chromatogram has to be inverted
by means of a derived channel. This means that the conductance measured is
multiplied by "-1" (see chapter 13).