HF scientific DRT-15CE Portable Turbidimeter User Manual

Page 11

Advertising
background image

DRT-15CE (5/10)

Page 3

4.0

OPERATION AND DESCRIPTION

To operate the turbidimeter, it is first necessary to standardize the instrument.

Switch to the “10” range and place the Calibration Standard (0.02 NTU) in

the optical well.

The EPA recommends that cuvettes used for instrument calibration or sample measurement

be indexed. For quick and repeatable indexing of the Calibration Standard, an indexing

ring and locator pin are included with this instrument.

When shipped, the white locator pin is installed in the collar ring around the

optical well of your turbidimeter. The indexing ring is included in the

accessory section of this instrument.

To index your Calibration Standard, slowly rotate the Calibration Standard, at

least one complete revolution, while observing the reading, and locate the

position of the lowest reading. Without moving the Calibration Standard,

install the indexing ring over the ridged cap of the Calibration Standard such

that the arrow on the o-ring aligns with the locator pin.

When indexing the Calibration Standard in the future, simply insert the

Calibration Standard and rotate it until the arrow on the o-ring faces the

locator pin. Please note that this Calibration Standard is only indexed to the

turbidimeter for which it was aligned.

To standardize, first index the Calibration Standard as above. Then adjust

the Reference Adjust in the appropriate direction to cause the display to read

0.02 NTU. The unit is now ready for use on any range.

To make a measurement of a sample, clean one of the cuvettes and fill it to

within approximately 1/2" (12 mm) of the top with the sample to be

measured. Place the cap on the cuvette and carefully clean the outside

surface of the cuvette with a lint free wiper such as "Kimwipes". Place the

sample in the well and take the NTU reading directly from display. Select the

appropriate range for best resolution..

If the instrument has been subjected to cold (below 10 degrees Celsius) and

then brought indoors, it should be allowed to warm up before use, since

condensation may form on the various lenses. Warm up can be aided by

leaving the case open and the instrument on for approximately a half hour.

Advertising