HF scientific DRT-15CE Portable Turbidimeter User Manual

Page 17

Advertising
background image

DRT-15CE (5/10)

Page 9

8.0

TROUBLE SHOOTING

Symptom

Meter does not respond when a sample is set into the well.

Possible Cause

1. Lamp is burned out. Replace the lamp and recalibrate.

2. Printed Circuit Board failure. Replace Printed Circuit Board.

3. Battery is low. Charge battery.

In the case of 1 or 2 the in strument should be recalibrated. The lamp has an

extraordinary long life and therefore replace ment is infrequent.

Symptom

Reference Adjust does not have enough travel to adjust for the calibration

standard value.

Possible Cause

1. Scratched or rubbed calibration standard container or aged calibration

standard. Replace the standard.

2. Optics have aged. Recalibrate.

3. Faulty lamp. Replace the lamp and recalibrate.

Symptom

The display will not stabilize with the calibration standard in the well.

Possible Cause

1. Battery voltage too low. "BAT" indicator is "ON" when the battery

requires recharging. When the battery is dis charged, the voltage will

drop off causing the meter display to drift. Recharge battery.

2. Cold sample can cause condensation on the cuvette. Clean and dry

cuvette.

3. Unit has not been given suf ficient time to stabilize at ambient

temperature. Allow time for instrument to stabilize.