HF scientific DRT-15CE Portable Turbidimeter User Manual

Page 14

Advertising
background image

DRT-15CE (5/10)

Page 6

Each kit contains:

--

Instruction manual

--

1 liter of 4000 NTU Stock Suspension

--

1 Gallon (3.79 liters) turbidity-free water

--

4 Sample cuvettes (28 mm)

--

4 Light Shield Caps

--

Graduated Pipettes 1 ea. in 1 ml, 10 ml, & 25 ml

--

1 Calibration Standard

The following table gives the recommended dilutions of the stock suspension.

Be sure to adequately mix the stock suspension prior to removing a portion

for dilution.

PROCEDURE

NTU Value

Pipette

Amount

Base

Formazin

Volumetric

Flask

10.0 NTU

10.0ml

2.5ml

of 4000 NTU

1000ml

100.0 NTU

10.0ml

5.0ml

of 4000 NTU

200ml

1000.0 NTU

25.0ml

25.0ml

of 4000 NTU

100ml

NOTE:

When the prepared samples start to flocculate, they are

unreliable and fresh ones must be made. This will occur

more rapidly for the lower value diluted suspen sions.

Prepare 40 NTU and lower standards daily.

7.2

ELECTRONIC CALIBRATION USING FACTORY

CERTIFIED PROCAL STANDARDS (CAT. NO. 39071)

The DRT Turbidimeters have been carefully calibrated at the factory.

However, should the Electronic P.C. Board, the Photo Detec tors, or the Light

Source be replaced recalibration may be easily accomplished at your facility.

To carry out a complete calibration the following ProCal Standard values are