HF scientific DRT-15CE Portable Turbidimeter User Manual

Page 12

Advertising
background image

DRT-15CE (5/10)

Page 4

5.0

RECORDER OUTPUT

The DRT-15CE has a 0-1 mA Recorder Output. The mA output jack is

located on right side of the chassis (refer to Figure 2). To use, connect the

1/8" miniplug provided to your recorder. Adjust R17 (pot nearest jack) to

obtain a full scale output compatible to a full scale reading on the DRT-

15CE. Once this adjustment is made, the DRT-15CE will always be set up

for this recorder. Use 10, 100 or 1000 NTU standard in appropriate range.

6.0

CRITICAL MEASURING AREA

The critical measuring area of the sample cuvette is a 3/4" wide band starting

5/8" above the bottom. Keep this area clean and free of scratches or

abrasion. Handle by the top part only. (See Figure 1).

Figure 1