C. SPECIFICATIONS FOR THE MICRO 200 BW

Specification

MICRO 200 BW

Ranges:

0 - 1000.0 NTU

Method:

EPA approved* Nephelometric

Accuracy:

10% 0-1000.0 NTU

Repeatability:

1%

Linearity:

+1%

Resolution:

0 to 10 NTU

0.0001

1 to 100 NTU

0.01

Digital Display:

LCD 6 digits

Analyzer Graphics:

Built in LCD, Graphics recorder

Clock Graphics:

Date and Time

Response Time:

1 to 8 seconds

Averaging Time (Electronic):

1, 16, 48 & 96 seconds, sliding average

Microprocessor:

Motorola MC68HC11

Keyboard data entry system:

8 interacting membrane switches with tactile feedback

Security Code:

Prevents unauthorized access to analyzer

Built in diagnostics:

Yes

Analog Output, Isolated:

Bracketed to .01 NTU, 4 - 20 mA & .1, 1 & 10 VDC 12 bit resolution D/A

Alarms:

1 system alarm, 2 (NTU or % T) alarms (Hi, Lo and Off)

Alarm Contact rating:

Max. 250 VAC @ 5.0 A

Operating Temperature:

32° to 122°F (0°C to 50°C)

Storage Temperature:

-4°F to 140°F (-20° to 60°C)

Optical Cleaning for NTU Sensor:

Removable cuvette

Flow rate:

0.5 - 1.5 gpm (2000 ml/min - 6.0 L/m)

Positive System Pressure:

60 psi maximum (414 kPa or 4.22 kg/cm

3

)

Wetted surfaces:

Nylon, C.A.B., PVC, Delrin-500, Borosilicate glass, Silicone

Lamp:

External, pre-focused module

Standard Cable Length:

6 feet (1.83 m)

Sensor to analyzer distance:

250 feet (76.25 m) max. Over 50' lengths contact factory.

Dimensions

Analyzer:

13" x 11½” x 8" (330 x 290 x 200 mm)

NTU Sensor:

8¾” x 9¾” x 8" (225 x 250 x 200 mm)

NTU Sensor Case:

NEMA 4X

CE Rating:

Pollution Degree 2, Overvoltage Category III

Supply Voltage:

120 VAC + 10% 50/60 Hz
240 VAC + 10% 50/60 Hz

Power Consumption:

40 VA

*EPA approved means that the instrument meets or exceeds the design and performance criteria as specified in the
United States Environmental Protection Agency Method 180.1.