b. Set Average

This is the Set Average Menu. This is where you can make changes to the averaging period. Readings are updated at a rate of one

per second. A higher averaging time will give smoother curves to the graphical display and to recording devices and helps “wash

out” anomalies. Always use the X1 setting when taking grab sample readings. Suppression or “Holding” the least significant digits

to zero may be set here. The X1 setting is automatically selected while in the calibration menu.

Selection Options:

F1:

Selects each reading (one second)

F2:

Selects Averaging of the last sixteen readings (16 seconds)

F3:

Selects Averaging of the last forty-eight readings (48 seconds)

F4:

Selects Averaging of the last ninety-six readings (96 seconds)

F5:

Selects between 1 digit, 2 digits & none of the least significant digit suppression

6. Utility Menu

This is the Utility Menu. The available sub-menu selections are listed below.

Selection Options:

F1:

Selects Screen Brightness & Clear History Menu (from graphs)

F2:

Selects Set Time & Date Menu

F3:

Selects Calibration Menu

F4:

Selects serial port & Access code set-up menu

F5:

Performs a check of internal critical voltages

F2

F3

ALARM

F5

F4

MENU

F1

Set Average

X1

XØ1

X16

X48

X96

Digit Suppress

Ø

F2

F3

ALARM

F5

F4

MENU

F1

Utility

Screen Brightness/History

Set Time

Calibrate

Comm Port

Self Test