HF scientific Micro200BW Turbidimeter - 0 -1000 NTU User Manual

Page 28

Advertising
background image

Page 14

MICRO 200 BW 0 - 1000 NTU (5/05)

Rev. 3.1

4. Alarm Menu

This is the ALARM MENU. The left side of the screen shows the selections as described below. The right side shows the current

settings. In the example above:

Alarm 1 is set to 1000 NTU and is off

Alarm 2 is set to 200 NTU and will alarm below 200 NTU

The Dead Band is set to 10%

Selection Options:

F1:

Not Used

F2:

Selects ALARM 1 setpoint MENU

F3:

Selects ALARM 2 setpoint MENU

F4:

Selects the DEAD BAND for the alarms to 2.5%, 5%, or 10%

F5:

Not Used

a. Alarm #1

This is the Alarm #1 Menu. In this menu you can set the value at which the alarm contacts for Alarm #1 will change. The purpose

of the function buttons are shown below. The current setting is displayed in the center top of the screen. Alarm #2 is displayed

and used exactly the same as this menu.

Selection Options:

F1:

Not Used

F2:

Selects the digit to change as indicated by the flashing cursor

F3:

Increments the selected digit (increases it)

F4:

Decrements the selected digit (decreases it)

F5:

Selects Alarm #1 to High, Low or Off

F2

F3

ALARM

F5

F4

MENU

F1

Alarm #1

Flashing cursor

F2

F3

ALARM

F5

F4

MENU

F1

Alarm Menu

Alarm 1

1ØØØ. Off

Alarm 2

Ш2ШШ. Lo

Dead Band

1Ш%

1ШШШ. OFF

Digit

Up

Down

Select Hi;Lo;Off