9.0 Troubleshooting/System Alarm Relay

9.1 System Warning Message(s)

The instrument routinely performs self-diagnostics and will automatically generate

warning messages to provide you with specific diagnostic information about the
instrument. These are only messages and they do not indicate failure of any component
in the instrument. Normally, the cause of a warning message is external of the
instrument. The following table lists the warning messages and their associated
meanings:

WARNING

ASSOCIATED MEANING

TYPICAL CAUSE

W01

Lamp failure

Lamp has too low an output or the 0.02

NTU calibration standard was

contaminated.

Attempt to recalibrate. If this does not

remedy the problem, the lamp must be

replaced followed by a recalibration.

W02

Calibration failure

The wrong standards were used in

calibration or there is an internal sensor

failure of the instrument.

W03

Analog loop failure

Analog loop connection is open. Verify

your wiring if you are using the analog

output. If not, turn off the option in the

setup menu.

W04

Flow Error

If the flow switch is option installed,

this indicates a loss of flow of process

to the instrument. If the flow switch is

not installed, turn off the option in the

setup menu.

O-r

Sample Over-Range

The sample turbidity level is greater

than the instrument will measure (either

11 NTU or 110 NTU depending on your

model).

WNG

Multiple Warning conditions are

met

A combination of warning conditions

has been met. Typically, this condition

arises when the setup for the analog

output and the flow switch are not set

correctly.


If any of the above conditions occurs, the message will be indicated on the lower row of

the display.

9.2 System Error Message
The instrument will also automatically generate error messages. An Error message is
generated when there is a significant problem found with the operation of the instrument.