Specifications – HF scientific 19944 Triple Validation Turbidimeter - 0-100 NTU User Manual

Page 6

Advertising
background image

TVT2 (5/02)
Rev. 1.7 (.pdf)

1

Specifications


Measurement Range

0.01 –100.0 NTU

Accuracy

±2% of reading plus 0.01 NTU (0-10 NTU)
±5% of reading plus 0.1 NTU (10-100 NTU)

Resolution

0.0001 NTU on low readings

Response Time

Less than 8 seconds

Display

Multi-Line Liquid Crystal Display

Two User Programmable
Alarms

120-240VAC 2A Form C Relay with suppressors

Analog Output Signals

4 - 20 mA

Water Pressure

0.1 - 8 bar (1-116 psi.)

Flow Rate

0.5 Liter/min. regulated (.26 Gal/min)

Operating Temperature Range 1°C – 50°C (34°F – 122°F)

Sample Temperature Range

1°C – 50°C (34°F – 122°F)

Power Supply Voltage

90 – 250, 47 – 90 Hz

Power Consumption

6 VA

Insulation Rating

Pollution Degree 2,
Overvoltage Category II

Regulatory Compliance (0-
10NTU)

White Light Version compliant to U.S. EPA 180.1
Infrared Version compliant to ISO 7027

Approximate Dimensions

175mm W x 310mm H x 225mm D (6.875” W x 12.25” H x
8.875” D)

Shipping Weight

2.5 kg (5.5 lbs.)

Warranty

1 Year from date of shipment