LaMotte DO 6 Dissolved Oxygen Meter User Manual

Page 34

Advertising
background image

7.5 (AtO.4) Automatic Off

This feature is useful for batteries energy
conservation. The meter automatically
shuts off 20 minutes from the last key
press.
From any Set Up Menu,
1. Scroll the Set Up main group menu

by using the ? and ? keys until the
meter displays the Automatic Off
menu “AtO.4”.

2. Press the HOLD/ENTER key to enter

the menu.

3. Use the

p and q keys to activate or

deactivate the automatic off selection.

a.

YES – activates Automatic Off

b.

NO – deactivates Automatic Off

4. Press the HOLD/ENTER key to

confirm the selection.

5. Press the CAL key to exit the Set Up menu to return to the measurement

mode.

7.6 (rSt.5) Reset to Factory Default

This mode lets you reset all parameters to factory default settings. There are
two levels of reset:
1. Calibration Reset: - Reset only the calibration values. This clears all

previous calibrated values and reset to factory default. See Section for
factory default settings.

2. User Reset: - Clears all data including calibration and other customized

setup functions and reset it to factory default. See Section 11 for factory
default settings.

33

HOLD

ENTER

Figure 24: Selection of Automatic

Off function