LaMotte DO 6 Dissolved Oxygen Meter User Manual

Page 48

Advertising
background image

% Saturation Calibration
Points

100% in saturated air or
air-saturated water.
0% in zero oxygen solution

% Saturation Calibration
Limits

Factory calibrated absolute value of
10.0% and below for 0% point &
50% to 200% for 100% point.

Concentration Calibration
Window

+/- 40% from the factory default
measurement value. Minimum
reading allowed is 2.00 mg/L (ppm).

Temperature Calibration
Window

+/- 5°C from factory default
measurement

Offset Adjustments (%
Saturation)

+/- 10.0 of reading in Saturation
mode

HOLD function

Yes

Auto-Off function

Selectable Auto Off function. (20
minutes after last key press)

Display

Customs Single 4 Digit LCD

Inputs

BNC for DO & 2.5 mm Phono for
temperature

Operating Range

0 to 50 °C

Power Requirements

4 AAA-sized batteries (included)

Battery Life

> 700 hours (Alkaline Batteries)

Dimensions

Meter: 14 cm (L) x 7 cm (W) x 3.5
cm (H)
Probe: 115 mm (L) x 12 mm (Dia)
with 3-ft cable.
Probe’s membrane housing: 16 mm
(Dia)