7.6.1 Calibration Reset
From any Set Up menu:
1. Scroll the Set Up main group menu by

using the

p and q keys until the meter

displays the Reset to Factory Default
menu “rSt.5”.

2. Press the HOLD/ENTER key to enter

the menu.

3. Press the HOLD/ENTER key to enter 4.

Use the

p and q keys to activate or

deactivate the Calibration Reset
selection.

a.

YES – activates Calibration Reset

b.

NO – deactivates Calibration Reset

5. Press HOLD/ENTER key to confirm the

selection made.

6. If selection is ‘YES’, meter would

immediately return to measurement
mode. Otherwise it will return to the
main group menu. Press CAL to return
to the measurement mode.

Note: When you reset your meter from %
Saturation Set Up menu, both % Saturation
and mg/L (ppm) Concentration calibrations
are reset to factory default. However, calibration reset in the mg/L (ppm)
Concentration Set Up menu will only reset the calibration in the
concentration mode and do not affect the calibration in the percentage
saturation mode.

.

34

HOLD

ENTER

HOLD

ENTER

Figure 25: Calibration reset

selection

Calibration
Reset