OR if Owner will retain Verifier, use this paragraph;

G.

Coordinate with owner retained verifier for final verification of the systems. Make corrections as needed,

including additional testing if necessary to illustrate full and unqualified certification.

COORDINATION

A.

Medical Gas Contractor shall coordinate with other trades to ensure timely installations and evade conflicts

and interference.

B.

Work with metal stud partition installer and/or mason to ensure anchors, sleeves and similar items are pro

vided in sufficient time to avoid delays; chases and openings are properly sized and prepared.

C.

Coordinate with owner to ensure medical gas outlets, whether owner supplied or contractor supplied, in

walls, ceilings and all equipments are provided by the same Medical Gas Equipment Manufacturer (MGEM)

are satisfactory to the owner.

D.

Coordinate with bulk cryogenic gas supplier for installation, connection and verification of bulk gas supply

systems.

E.

Medical Gas Contractor shall supply and install the master alarm system (including the signal wiring). The

electrical contractor shall provide power wiring to each alarm panel. Medical Gas Contractor is accountable

for proper termination, testing and marking of alarm panels. Termination shall be done by or under the

supervision of the manufacturer of alarm panels.

F.

Coordinate with Medical Gas Verifier to deliver a complete, tested medical gas installation ready for owner’s

use.

SUBMITTALS

A.

Furnish the following as one package:

1.

Medical Gas Equipment Manufacturer (MGEM) submittals shall include at least or at a minimum of

the

following:

a.

Complete specifications for the product intended to be installed, dimensional drawings,

and wiring schematics where appropriate.

b.

For other medical gas products include:

i.

Package drawing indicating package style, dimensions when complete, method of

disassembly and sizes of subsections for rigging and installation.

ii.

Compressor and package capacity expressed in L/min along with SCFM.

iii.

Lubrication method (if any).

iv.

Drive detail including adjustment method.

v.

Motor including frame type, service factor, kilowatts, current draw, and RPM.