Latch valve assembly puritan-bennett (compatible) – Amico Retractable Ceiling Column Outlet User Manual
Page 11
Latch Valve Assembly Puritan-Bennett (Compatible)
19
20
17
11
14
15
12
18
16
USE N
O OIL
P
R
E
S
S
T
O
R
E
LE
S
E
P
R
E
S
S
T
O
R
E
LE
S
E
1. Unscrew the two retaining screws (12) until the Latch Valve Assembly (11) can be removed from the outlet.
2. Remove the two screws (20), holding the round retaining ring (19) in place. Remove the plate.
3. Remove the gas connector (18) from the valve assembly.
4. Remove the Body Seal (15) and Poppet (16) from the front of the Gas Connector (18). Inspect the items for wear or
damage and replace the Body Seal (15).
5. Remove the Gas or Vacuum Spring (17), to inspect for wear or damage. Re-install the Gas or Vacuum Spring.
6. Re-install all internal components into the Gas Connector (18). Check that the Gas or Vacuum Spring (17), Body Seal
(15) and Poppet (16) are in place. Re-install the round retaining plate (19) and secure with two screws (20), do not
overtighten.
7. Re-install the Latch Valve Assembly into the outlet. Coat the connector (18) with a thin coat of oxygen compatible
silicone lubricant to aid insertion. Tighten down the retaining screws (12), DO NOT over tighten, as this could
damage the Latch Valve.
8. Connect a gas specific adapter into the outlet. The connection should be smooth and the adapter should lock and
remain in place allowing gas to flow. If not, replace the entire Latch Valve Assembly (11).
Note:
Latch-Valve Assembly Model
Numbers see Page 14
Complete Outlet Model Numbers
see Page 16
Repair Kit Model Numbers see
Page 21
www.amico.com
11