Installation and dimensions – Amico Retractable Ceiling Column Outlet User Manual

Page 5

Advertising
background image

Installation and Dimensions

Wall thickness

may vary from

10mm to 25mm]

3/8" to 1"

[

1/2" [12.7 mm] O.D. (3/8” nominal) type "K"

copper pipe, rotates 360 deg.

for entry from any angle

5-1/2”

[140]

NOTE:

Amico recomends that you have a blank space

between outlets and slides.

Maximum 3 outlets may be ganged together

without additional support.

5-27/32” [ 148]

3” [ 76]

5”

[127]

5-3/4”

[146]

10”

[254]

1-1/16”

[27]

3/8”

[10]

5”

[127]

5”

[127]

5”

[127]

1 7/8”

[48 ]

OXYGEN

YGEN

OXY

OXY

When wall outlets are ganged
together, ensure that they
are 5” from center to center.
DO NOT let the notches on
the mounting plate pass the
adjacent holes on the next
mounting plate.

Do not bury the outlet during
the drywall and plaster appli-
cation. Ensure that the cover
over the outlet is in place un-
til the “Latch-valve assembly”
can be installed.

Wall Outlets

DISS, Ohmeda compatible, Chemetron compatible, Puritan-Bennett and Oxequip/MedStar compatible

Inch

[mm]

www.amico.com

5

Advertising