Installation and dimensions, Inch [mm, 6amico pipeline – Amico Retractable Ceiling Column Outlet User Manual
Page 6
Wall thickness
may vary from
[10mm to 25 mm]
3/8” to 1”
5”
[127]
5”
[127]
5”
[127]
NOTE:
Amico recommends that you have a blank space
between outlets and slides.
Maximum 3 outlets may be ganged together
without additional support.
5-1/2”
[140]
3”
[76]
5-27/32”
[148]
5”
[127]
5-3/4”
[146]
1-1/4”
[32]
3/8”
[10]
1-7/8”
[48]
MEDVAC
Ceiling Outlets
DISS, Ohmeda compatible, Chemetron compatible, Puritan-Bennett and Oxequip/MedStar compatible
When wall outlets are ganged
together, ensure that they
are 5” from center to center.
DO NOT let the notches on
the mounting plate pass the
adjacent holes on the next
mounting plate.
Do not bury the outlet during
the drywall and plaster appli-
cation. Ensure that the cover
over the outlet is in place un-
til the “Latch-valve assembly”
can be installed.
Inch
[mm]
Console Outlets
DISS, Ohmeda compatible, Chemetron compatible, Puritan-Bennett and Oxequip/MedStar compatible
TOP VIEW
1/2" [12.7 mm] O.D. (3/8” nominal) type "K"
copper pipe, rotates 360 deg.
for entry from any angle.
SIDE VIEW
FRONT VIEW
BACK
ASSEMBLY
FRONT VIEW
FRONT
ASSEMBLY
SIDE VIEW
FRONT
ASSEMBLY
2-7/16"
[62]
3-3/4"
[95]
2-7/16"
[62]
1-13/16"
[46]
3-9/32"
[83]
3-9/16"
[91]
8-7/8"
[225]
OXYGEN
OXY
OXY
inch
[mm]
Installation and Dimensions
6
Amico Pipeline