Setting Up a Custom Application

User Guide | 79

3.

Enter a unique application name.

4.

Select a filter, scan mode, and image color from the lists.

When selecting the scan mode, you have a choice:

Transmissive — light from the top of the densitometer passes through the

gel. Choose this option if you are scanning gels or transparent objects,

such as X-rays or film.

Reflective — light from the bottom of the densitometer reflects off the gel.

Choose this option if you are scanning blots or oblique objects, such as

paper or photographs.

5.

Click OK to create the custom application.