Bio-Rad MicroRotofor™ Cell User Manual

Page 15

Advertising
background image

2. Remove the two screws that secure the Harvesting station to the vacuum

block and remove the positioning block, needle array, and needle array
holder (Figure 15).

Fig. 15. Disassembling the harvesting station (top) and harvesting station components
(bottom).

3. Clean and dry the positioning block.

4. Using a wash bottle, wash and rinse the individual needles in the needle

array.

5. Clean and dry the vacuum chamber.

6. Inspect and reposition the vacuum gaskets.

7. Reassemble the harvesting station.

13