1.2 Specifications

Construction

Focusing chambers

Acrylic

Vent buttons

Porous polytetrafluoroethylene (PTFE)
membrane in molded plastic

Gaskets

Silicone rubber

O-Rings

Fluorocarbon elastomer

Cooling finger

Ceramic

Housing

Polycarbonate and acrylic

Harvest box and lid

Polycarbonate and acrylic

Tubing

Polyvinyl

Needle array

Stainless steel and acrylic

Electrodes

Platinum, 0.010 inch diameter

Membrane Core

Molded polyethylene with polyester mem-
branes

Chemical

The Rotofor cell components are not com-

compatibility

patible with chlorinated hydrocarbons
(

e.g. chloroform), aromatic hydrocarbons

(

e.g. toluene, benzene), or acetone.

Use of organic solvents voids all warranties.

Shipping weight

9 kg

Overall size

45.7 cm (L) x 16.5 cm (W) x 22.8 cm (H)

Cell voltage limit

3000 VDC

Cell power limit

15 W

Cooling

The Rotofor cell must be run with cooling or
excessive heating may occur, damaging the
unit. A refrigerated circulating water bath is
recommended to keep the coolant temperature at
4 °C.

Maximum coolant

12 L/minute

flow rate
Minimum coolant

50 ml/minute

flow rate
Sample volume

18–58 ml

Electrical

3 wire cord

connection
Input power

120 V Model: 100-120 VAC, 50/60 Hz, 12W

Requirements

240 V Model: 220-240 VAC, 50/60 Hz, 12W

Fuses

250 mA Type T (1 required, 1 spare)