Error indicators accessories – Iris Sample Processing StatSpin® CytoFuge 12 Versatile Cytocentrifuge System User Manual

Page 10

Advertising
background image

5

Error Indicators

Cytofuge

®

12 Cytocentrifuge error conditions are signaled by flashing codes appearing in the TIME

display on the front panel.

To clear an error condition, correct the problem and press the Stop/Open button.

Error Codes

Acceleration Error
Confirm no obstruction in bowl area, check
drive cone o-rings (see Section 5), contact
customer service.

Latch is not locked
Confirm cover is fully seated in closed
position.

Incorrect Speed
Remove and inspect rotor, check drive
cone o-rings (see Section 5), contact
customer service.

Cover is not closed
Confirm cover is fully seated in closed
position.

Out of balance
Confirm concentrators are balanced in
rotor and cytocentrifuge is on a level
surface.

Accessories

Product No.

Description

CC03

3-well Cell Concentrators and 3-well gaskets; Box of 10

CC01

1-well Cell Concentrators and 1-well gaskets; Box of 10

CLIP

Stainless Steel Clips for all Cell Concentrators; Bag of 10

GK01

Replacement Gasket for CC01 and CCR1

; Box of 100

GK03

Replacement Gasket for CC03 and CCR3

; Box of 100

CCR1

1-well Cell Concentrators, reusable, autoclavable; Box of 10

CCR3

3-well Cell Concentrators, reusable, autoclavable; Box of 10

BP01

Backing Plates for all Cell Concentrators; Box of 10

FF01

Filter Concentrators , Disposable; Box of 48

FF01-B

Filter Concentrators , Disposable; Box of 200 Filter Cards, 192 Concentrators

FFR1

Filter Concentrators, Reusable; Box of 200 Filter Cards, 20 Concentrators

FFCL

Stainless Steel Clips for Filter Concentrators; Box of 4

CFRT-12

Cytofuge 12 Rotor Assembly, replacement or spare, includes cover

CFRC-12

Cytofuge 12 Rotor Lid, replacement