How to use this manual – Iris Sample Processing StatSpin® CytoFuge 12 Versatile Cytocentrifuge System User Manual

Page 4

Advertising
background image

ii

How to use this manual

This manual along with information contained on product labels and in package inserts should provide
you with all the information you need to operate and maintain the Cytofuge 12 Cytocentrifuge system.

When the information requires special attention, a caution symbol appears next to the italicized
text.

Please pay close attention to the instructions that accompany the notes and symbols as well as the
standard laboratory practices outlined by your facility and local regulatory agencies. The table below lists
all the CAUTIONS and WARNINGS for the Cytofuge

®

12 Cytocentrifuge.


WARNING - Do not expose the cytocentrifuge or its rotor to strong or concentrated acids, bases or
oxidizing agents, aromatic or halogenated hydrocarbons, esters or ketones.

WARNING - Do not use the supplied line cord outside of North America. Use a power cord with a
grounded male connector suitable for the power outlet being used, which is rated at 3.0 Amps or more
and has an IEC320/CEE22 female connector to attach to the cytocentrifuge.

WARNING - Maintain a 300 mm (11.8 inches) clearance envelope around cytocentrifuge during
operation.

Do not deposit any potentially hazardous materials within the CLEARANCE ENVELOPE.

WARNING - Inspect instrument for cracks or any physical damage to housing, cover, lid and rotor.
Damage may result in unsafe operation; discontinue use until repairs have been performed.

CAUTION - Do not overfill the concentrators. Addition of excess amounts of liquid to Cell or Filter
Concentrators will result in the fluid being “spun out” during centrifugation.

CAUTION - Do not place the cytocentrifuge on top of other equipment and do not place other objects or
equipment on top of the cytocentrifuge.

CAUTION - The cover interlock by-pass is for emergency use only. Do not attempt to operate the
Cytofuge 12 while the latch release tool is inserted in the housing. Misuse of the interlock bypass may
expose the operator to unsafe conditions.

CAUTION - Before using any cleaning or decontamination methods except those recommended,
operators should check with customer service that the proposed method will not damage the
equipment.

CAUTION - The rotor (CFRT-12) is designed for many years of service but, depending on the
frequency of use, the rotor components may develop cracks over time or if dropped on a hard surface.
Inspect the rotor for damage or cracks before each use. If damage is found discontinue use and
contact customer service or your distributor for a replacement.

CAUTION - Do not lift the rotor (CFRT-12) by the locking button - the lid may release allowing the rotor
bowl to fall. Always lift the rotor by the black center grip.

CAUTION - The rotor (CFRT-12) is intended for use with biological samples only. Do Not place
flammable or reactive substances in the rotor.

CAUTION - While in the resting position, the rotor is designed to maintain the Concentrators at an
angle to prevent fluid from contacting the slide and/or the filter card prior to spinning. Keep the rotor
level during transport to and from the cytocentrifuge to ensure proper concentrator angle.

CAUTION - Do not defeat any safety interlocks while checking the rotor speed. Bypassing interlocks
could result in personal injury.

CAUTION

-

Unplug the cytocentrifuge from the wall outlet before cleaning or performing maintenance.