Specifications – ITC Hemochron Signature+ Whole Blood Microcoagulation System User Manual

Page 10

Advertising
background image

10

SPECIFICATIONS
Specifications for the HEMOCHRON Jr. Signature+ Whole Blood Microcoagulation System are listed below.

Dimensions and Weight

Depth

9 cm (3.5 in)

Width

19 cm (7.5 in)

Height

5 cm (2.0 in)

Weight

0.60 kg (1.3 lbs)

Performance

Test Precision

≤ 10% C.V.

Operation

Test Chamber

1

Timing Range

0 second to 1005 seconds

Incubation Temperature

37 °C ±1.0 °C

Incubation Warm-Up Time

30 seconds to 90 seconds

Full-Charge Operating Time

2 to 3 hours (minimum)

Battery Life

500 recharges

Battery Type

Nickel Metal Hydride

Throughput (Full Charge)

49 test cycles (at 150 sec per test)
17 test cycles (> 500 sec per test)

Operating Environment

15 °C to 30 °C

AC/DC Power Module

USA/Canada

Input

Power

120 VAC, 60 Hz, 35 VA

Output

Power

9.0 VDC, 1.38 Amps maximum

Europe

Input

Power

230 VAC, 50 Hz, 30 VA

Output

Power

9.0 VDC, 1.15 Amps maximum

Japan

Input

Power

100 VAC, 50/60 Hz, 40 VA

Output

Power

9.0 VDC, 1.3 Amps maximum

Calibration
The Signature+ instrument is calibrated at the manufacturer facility to test and verify all functions. The
instrument is also self-calibrating, as all instrument functions are continuously monitored and verified by
the instrument software when a test is performed. The instrument does not require additional calibration
by the user.