To Unlock an Instrument that has Reached QC Failure Limit or the 911 Limit:
Note
: The HEMOCHRON Jr. Signature+ must be connected to the Configuration Manager and
the instrument configuration information read by clicking on Read Configuration from

Instrument.

1.

Select Lock-Out Control from the Configuration Module screen. The EQC/LQC Lockout dialog
box is displayed.

2.

Use the drop-down arrow to select the test (e.g., PT or EQC) for which QC failed. The failed QC
Normal and QC Abnormal tests are displayed in the #of QC Tests Failed fields.

3.

Click Reset for QC Normal or QC Abnormal to reset the number of failed QC tests to zero for that
level.

4.

Select the QC Exceptions tab and click Reset to reset the number of used patient tests after
EQC/LQC Lock-Out.

5.

Click Save & Exit. The Configuration Module dialog box is displayed (see page 31).

Note: The reset configuration must be downloaded to the instrument for these changes to
take effect.

Specifying Operator IDs and/or Personal Identification Numbers
The Operator Table allows the supervisor to specify an Operator ID (OID) and/or Personal Identification
Numbers (PINs) for each operator or a group of operators. The supervisor can then specify whether entry of
a specified OID or PIN is required before the instrument can be operated.

1.

Select Operator Table from the Configuration Module screen. The Operator Table dialog box is
displayed:

2.

Enter a numeric ID for each operator. The ID can be any number (up to 9 digits) other than
zero (0).

3.

Enter a numeric PIN for each operator. The PIN can be any number (up to 9 digits) other than
zero (0).

Note: A unique PIN can be assigned for each authorized operator, or a single OID and PIN
can be used to allow operation by any operator that enters the correct code.

4.

The Status box will display the OID/PIN requirements set under Instrument Options. These
options are all mutually exclusive.

5.

Click Save & Exit to save changes to the Operator Table and exit the dialog box.

Note: Cancel removes the dialog box without saving changes. Reset clears all information
from the Operator Table.