17

Entering a Mandatory PID, OID, and/or PIN
Entry of a Patient ID (PID), Operator ID (OID), and/or Operator Personal Identification Number (PIN) can
be required for completion of a test (see the Configuration Manager section on page 26). If entry of one or
more of these IDs is required, appropriate prompts for entry of PID, OID and/or PIN are automatically
displayed.

To Enter a Mandatory Patient ID:

1.

The PID prompt is automatically displayed:

PID= . . . . . . . . . . . .

Note: When PID=… is displayed, pressing the QC button will allow a QC test to be
performed without fulfilling the PID requirement.

2.

Type the PID using the numeric keys.

3.

Press ENTER. If an acceptable PID is entered, operation of the instrument continues:

Priming Pump

Note: An acceptable PID is any number other than zero (0). The number of digits to be
entered for PID (from 3 to 9, or 0 for any number of digits up to 9) may be designated via
Configuration Manager. Selecting 3 to 9 digits requires that the operator enter exactly that
length. Check the PID by pressing ID and then pressing 1 to display the entered value.

To Enter a Mandatory Operator ID:

1.

The OID prompt is automatically displayed:

OID= . . . . . . . . . . . .

2.

Type the OID using the numeric keys.

3.

Press ENTER. If an acceptable OID is entered, operation of the instrument continues:

Priming Pump

Note: If entry of any OID is specified, an acceptable OID is any number (up to 9 digits) other
than zero (0). If entry of a valid OID is specified, an acceptable OID is any number contained
in the Operator Table (see page 36 ).
Check the OID by pressing ID and then pressing 2 to display the entered value.

4.

If an unacceptable OID is entered, ID IS NOT VALID is displayed followed by the OID prompt.
Operation of the instrument will not continue until an acceptable OID is entered.

To Enter a Mandatory Operator PIN:

1.

The PIN prompt is automatically displayed:

PIN= . . . . . . . . . . . .

2.

Type the Operator PIN using the numeric keys.

Note: The Operator PIN will be displayed as ** on the display panel. The corresponding OID
will be stored and printed with results.

3.

Press ENTER. If an acceptable PIN is entered, operation of the instrument continues:

Priming Pump

Note: An acceptable PIN is any number contained in the Operator Table.

4.

If an unacceptable PIN is entered, ID IS NOT VALID is displayed followed by the PIN prompt.
Operation of the instrument will not continue until an acceptable PIN is entered.