HI-4450 Isotropic Electric Field Probe

© ETS-LINDGREN L.P. – December 2004
REV B - PN H-600086A

6. APPENDIX A:

HI-4450/FP2080/FP4080/FP5080
ERROR CODES

PROBE ERROR OUTPUT

If an error occurs, the probe will respond with one of the
following strings. These strings begin with a colon and end with
a carriage return.

E01

Communication error (e.g., overflow).


E02

Buffer full error. Too many characters contained

between the Start Character/Carriage Return sequence.

E03

The received command is not valid.


E04

The received parameter is not valid.


E05

Hardware error (e.g., EEPROM failure).


E06 Parity

error.