Guidance and manufacture’s declaration – electromagnetic immunity –

for all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic immunity

The pro6

+

is intended for use in the electromagnetic environment specified below. The customer or user of the

pro6

+

should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Electrostatic

±6 kV contact

±6 kV contact

Floors should be wood, concrete or

discharge (ESD)

±8 kV air

±8 kV air

ceramic tile. If floor are covered with

IEC 61000-4-2

synthetic material, the relative humidity

should be at least 30%.

Electrical fast

±2 kV for power

±2kV for power

Main power quality should be that of

transient/burst

supply lines

supply lines

a typical commercial or hospital IEC

61000-4-4

environment.

Surge

± 1 kV line(s) t

±1 kV differential

Main power quality should be that of

IEC 61000-4-5

o line(s)

mode

a typical commercial or hospital

environment.

Voltage dips, short

<5% U

T

<5% U

T

Main power quality should be that

interruptions and

(>95% dip in U

T

)

(>95% dip in U

T

)

of a typical commercial or hospital

voltage variations

for 0.5 cycle

for 0.5 cycle

environment. If the user of the pro6

+

on power supply

requires continued operation during

input lines

40% U

T

40% U

T

power main interruptions, it is IEC 61000-

4-11

(60% dip in U

T

)

(60% dip in U

T

)

recommended the pro6

+

be powered

for 5 cycles

for 5 cycles

from an uninterruptible power supply or a

battery.

70% U

T

70% U

T

(30% dip in U

T

)

(30% dip in U

T

)

for 25 cycles

for 25 cycles

<5% U

T

<5% U

T

(>95% dip in U

T

)

(>95% dip in U

T

)

for 5 sec

for 5 sec

Power frequency

3A/m

3A/m

Power frequency magnetic fields should

(50Hz) magnetic

be at levels characteristic of a typical

61000-4-8

location in a typical commercial or

hospital environment.

NOTE U

T

is the a.c. main voltage prior to application of the test level.