Guidance and manufacture’s declaration – electromagnetic immunity –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacture’s declaration – electromagnetic immunity

The

pro6

+

is intended for use in the electromagnetic environment specified below. The customer or user of the

pro6

+

should assure that it is used in such an environment.

Immunity test

IEC 60601 test

Compliance

Electromagnetic environment

level

level -

guidance

Portable and mobile RF communications

equipment should be used no closer to

any part of the pro6

+

, including cables, than

the recommended separation distance

calculated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance

Conducted RF

3 Vrms

3 Vrms

IEC 61000-4-6

150 kHz to 80 MHz

Radiated RF

3 V/m

3 Vrms

80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

800 MHz to 2.5 GHz

Where

P is the maximum output power rating

of the transmitter in watts (W) according

to the transmitter manufacturer and

d is the

recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey,

a

should be less than the compliance level in each

frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following symbol:



NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the pro6

+

is used exceeds

the applicable RF compliance level above, the pro6

+

should be observed to verify normal operation. If abnormal

performance is observed, additional measures may be necessary, such as reorienting or relocating the pro6

+

.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

P

d

167

.

1

=

P

d

167

.

1

=

P

d

167

.

1

=