5

Use the

left

navigation key to select the A/D, then press the center softkey,

ADD.

The A/D is then moved to the

associated A/D’s column as shown in example 2. If no other A/D’s are listed in the

AVAILABLE A/D’s

column, the

center softkey changes to

Done

, as shown in Figure 8. Press

Done

to exit the Scale Configuration display.

Figure 8 Scale Configuration Displays

Analog Input Configuration

Use the following steps to work down to the ALGIN menu and fully configure the analog input option for the

920i

.

Figure 9 Analog Input Menu

1. Select SCALES from the main menu.

2. Scroll down and select the desired scale input.

3. Select SCALE 1 to set scale 1 configuration.

4. Select ALGIN to select the desired input ranges:

Temp

20 mA

10 V

100 mV

Therm

5. Press the

Enter

key to temporarily store a highlighted setting.

NOTES:

Settings are saved upon pressing the

Save

and

Exit

softkeys upon completion of the configuration and calibration

process.

The Restore softkey restores the highlighted input parameter to its previously saved setting. The Default softkey restores
the highlighted input parameter to its default setting.

Thermocouple Type Setup

There are five types of thermocouple probes that are supported by the

920i

. The probe types include Types E, J, K,

N, and T. They are all handled internally by

920i

core software, Version 2.05 and later.

Example 1

Example 2