Calibration, Parts kit contents – Rice Lake 920i Dual Channel Analog Input Card User Manual

Page 7

Advertising
background image

7

Calibration

To calibrate the indicator using the front panel, use the following steps.

1. Place the indicator in setup mode (display reads

Scale Configuration

).

2. With the Scales menu highlighted, press the

down

arrow key, then select the scale to be calibrated. Press

down

again, then press

left

to highlight the

CALIBR

sub menu.

3. Use

down

arrow to go to zero calibration (

WZERO

).

4. Set input current or voltage to the desired value.

5. Use the

down

arrow key to view the previous calibration value.

6. With the input value set to the lowest desired level, press the

Calibrate

softkey.

7. Use the

up

and

right

arrow keys to go to

WVAL

.

8. Use the

down

arrow key to view the previous

WVAL

value.

9. Enter the display value proportional to the calibration input signal (

WSPAN

).

10.Press

enter

.

11. Use the

up

and

right

arrow keys to go to

WSPAN

.

12.Adjust the input to the desired input value.

13.Use the

down

arrow key to view the previous value.

14.Press the

Calibrate

softkey.

15.Use the

up

arrow key to return to

SCALES

.

16.Press

Save

and

Exit

to return to normal operating mode.

Parts Kit Contents

Table 4 lists the part kit contents for the analog input option.

Specifications

A/D Resolution

14 Bit Effective Resolution

Available Channels

2

PN

Description

14626

Kep nuts, 8-32 NC (2)

14822

Machine screw, 4-40NC x 1/4 (2)

15133

Lock washers, NO 8 type (2)

15631

Cable tie (3-inch nylon) (4)

15885

4-position terminal block (1)

53075

Cable shield ground clamp (2)

73997

Multiple cable bushing (1)

76513

4-position screw terminal connector (2)

85357

Dual analog input card (1)

87702

Sealing plug, 3 x 9 mm (1)

Table 4 Parts Kit Contents

Range

Sensitivity @ 60 Hz

+/– 10V

+/– 1 mV

+/– 100 mV

+/– .05 mV

4-20 mA / 0020 mA

+/– .05 mA

Table 5 Analog Input Ranges