Calibration switch, How to calibrate the scale – Rice Lake Intercomp CS 1500 - Users Manual User Manual

Page 17

Advertising
background image

CS1500, Users

Rev S, June, 2009

17

Calibration switch

The calibration of the scale is protected from accidental change by a shunt placed on
“CAL” (instead of “RUN”) of J8, located on the top/middle of the A/D 20 BIT circuit
board.

Enabling the Calibration
The shunt must be moved from “RUN” to “CAL”. When you are done calibrating,
place the shunt back on “RUN” to protect the calibration from change.

How to calibrate the scale

The following details the calibration procedure for the crane scale. There are eight
parameters that can be set without moving the calibration blocking switch, followed by
five more parameters and calibration that require the calibration blocking switch be in
the enabled position.

Step

Display Parameter

Note

Default

4

ee-ee

ee-ee

ee-ee

ee-ee Skip

0=no skip, 1=EE-10

EE-10

EE-10

EE-10, 2=LL-00

LL-00

LL-00

LL-00

0

5

EE-00

EE-00

EE-00

EE-00 Sample Rate

1 to 64

4

6

EE-01

EE-01

EE-01

EE-01 Update Rate

1 to 32

4

7

EE-02

EE-02

EE-02

EE-02 Demand Output

1=Yes, 0=Continuous

1

8

EE-03

EE-03

EE-03

EE-03 Baud Rate

0 to 9

0

9

EE-04

EE-04

EE-04

EE-04 Auto-off Time

0 to 255; 0=off

20

10

EE-05

EE-05

EE-05

EE-05 Power up in KG

1=kg, 0=lb

0

11

EE-06

EE-06

EE-06

EE-06 Tare Type

1=keyboard, 0=push-button

0

12

EE-07

EE-07

EE-07

EE-07 Sleep Mode Time

0 to 255; 0=off

5

13

ee-08

ee-08

ee-08

ee-08 Set Point 1

1 to 99999

99999

14

ee-09

ee-09

ee-09

ee-09 Set Point 2

1 to 99999

99999

Information saved

Check for calibration blocking switch

15

EE-10

EE-10

EE-10

EE-10 AZT

0=off,1=0.6,2=1,3=3

2

16

EE-11

EE-11

EE-11

EE-11 Zero Range

0=off,1=on

0

17

EE-12

EE-12

EE-12

EE-12 Canadian Specifications 0=off, 1=on

0

18

EE-13

EE-13

EE-13

EE-13 Initial Zero Range

0=off, 1=on

0

19

EE-14

EE-14

EE-14

EE-14 Graduation

0 to 11

6

20

Information saved

21

LL-00

LL-00

LL-00

LL-00 Zero read

Enter capacity

22

LL-01

LL-01

LL-01

LL-01 First weight

Enter first weight

23

LL-02

LL-02

LL-02

LL-02 Second weight

Enter second weight

24

LL-03

LL-03

LL-03

LL-03 Third weight

Enter third weight

25

Information saved

Start up
1. Move the calibration blocking switch to the “CAL” position if you intend to

calibrate.