Rice Lake Intercomp CS 1500 - Users Manual User Manual

Page 18

Advertising
background image

CS1500, Users

Rev S, June, 2009

18

2. Turn scale power ON and wait for scale to warm up (3 minutes from power

on).

3. Press ZERO and lb/kg keys together and release to enter the calibration

mode.

First ten parameters
4. The scale shows “EE-EE

EE-EE

EE-EE

EE-EE”. Simultaneously press the ZERO and lb/kg keys.

To skip to “EE-10

EE-10

EE-10

EE-10” enter a “1”. To skip to “L

LL

LL-00

L-00

L-00

L-00” enter “2”. No skips will

occur with an entry of “0”.

Note: If you don’t know how to enter a number, see section titled “How to
enter a number”.

5. The scale shows “EE-00

EE-00

EE-00

EE-00”. Simultaneously press the ZERO and lb/kg keys.

Enter the sample rate (1 to 64). The sample rate is the number of past
readings that are averaged together to make a reading.

6. The scale shows “EE-01

EE-01

EE-01

EE-01”. Simultaneously press the ZERO and lb/kg keys.

Enter the update rate (1 to 32). The update rate is the speed at which the
displayed weight is updated. The smaller the number the faster the display
will be updated.

7. The scale shows “EE-02

EE-02

EE-02

EE-02”. Simultaneously press the ZERO and lb/kg keys.

Enter demand versus continuous on the optional serial output. See Serial
Output section.

Setting

Type

0

Continuous

1

Demand

8. The scale shows “EE-03

EE-03

EE-03

EE-03”. Simultaneously press the ZERO and lb/kg keys.

Enter the baud rate of the serial output (0 to 7).

Setting

Baud Rate

0

9600

1

4800

2

2400

3

1200

4

600

5

300

6

150

7

75

8

19.2K

9

38.4K


Advertising