CS1500, Users

Rev S, June, 2009

21

3: The push-button zero and AZSM can only operate within +/-

2% of the IZSM.

4: The over-capacity point is 103% of capacity above the IZSM.


18. The scale shows “EE-13

EE-13

EE-13

EE-13”. Simultaneously press the ZERO and lb/kg keys.

The scale shows the current initial zero range setting. Use the following table
to select the “initial zero range” setting:

Setting

Initial zero range

0

Off

1

On

A “0” setting implies:

1: The initial zero setting mechanism (IZSM) will work over the

entire range of the scale capacity.

2: The over-capacity point is 103% above the zero obtained at

calibration.

A “1” setting implies:

1: The IZSM must be within +/- 10% of the zero obtained at

calibration.

2: The over-capacity point is 103% above the IZSM.

19. The scale shows”EE-14

EE-14

EE-14

EE-14”. Simultaneously press the ZERO and lb/kg keys.

The scale shows the current graduation selection. Use the following table to
select a graduation value.

Settings

Count by in lb

Count by in kg

0

100

50

1

50

20

2

20

10

3

10

5

4

5

2

5

2

1

6

1

0.5

7

0.5

0.2

8

0.2

0.1

9

0.1

0.05

10

0.05

0.02

11

0.02

0.01

12

0.01

0.01

NOTE: The stated accuracy specifications are based on the graduation setting in
the table below. If the graduation setting is set other than the value in the table
below the accuracy specification remains with the graduation size listed below.