M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L

MSI-4300 Porta-Weigh Plus/5 Key

User Guide Page 13

CALIBRATE SETUP MENU

Top Menu

Selections

FILTER

LOW, MED, HIGH

UNITS

kg, lb, Ton, Metric Ton

STD CAL

Starts Cal procedure

FINE CAL

For minor calibration adjustments

CAL RCAL

For emergency calibration only

STANDARD

INDUSTRY, NIST, OIML, ALASKA, CANADA, UNITED

KINGDOM, LEGAL-FOR-TRADE, AUSTRALIA

AUTO OFF

ENABLED, DISABLED

AZM

ENABLED, DISABLE

REMOTE

ENABLED, DISABLED

LIGHTING

AUTOMATIC, ON, OFF

CONTRAST

(eight steps with numerics at “123456” and message reading

“CONTRAST”)

EXIT

Push ENTER (TEST) to quit setup

HI SET PT

HI VALUE

DISABLED, GROSS, NETGROSS

RESPONSE

NORMAL, BLANKLCD, BLINKLCD

EXIT

LO SET PT

LO VALUE

INPUT WT

MODE

DISABLED, GROSS, NETGROSS

RESPONSE

NORMAL, BLANKLCD, BLINKLCD

EXIT

TOTALIZE

MANUAL, AUTO

STANDARD CALIBRATION

Standard Calibration is to be used for normal test weight or cal fixture calibra-
tion of the Porta-Weigh Plus.

1) Enable Calibration by pushing the key behind the seal port.
2) Use the SELECT (NET/GROSS) key to scroll to the “STD CAL”

message. Push ENTER (TEST).

3) The message reads “Ø SCALE”. Remove all weight from the scale.

When motion ceases push ZERO. Assuming the detected zero weight is
within acceptable limits the message reads “WEIGHT” indicating it is